STUDY OF MATRIX EFFECT ON THE ANALYSIS OF Ni AND Pd BY AAS IN THE DESTRUATS OF HIDROCRACKING CATALYSTS USING AQUA REGIA AND H2SO4

https://doi.org/10.22146/ijc.21914

Wega Trisunaryanti(1*), Mudasir Mudasir(2), Siti Saroh(3)

(1) Chemistry Department, Faculty of Mathematics and Natural Sciences Gadjah Mada University, Yogyakarta
(2) Chemistry Department, Faculty of Mathematics and Natural Sciences Gadjah Mada University, Yogyakarta
(3) Chemistry Department, Faculty of Mathematics and Natural Sciences Gadjah Mada University, Yogyakarta
(*) Corresponding Author

Abstract


 The matrix effect of various metals on the analysis of nickel and palladium in destruate of catalysts of Butonian asphalt-derived asphaltene by atomic absorption spectroscopy have been investigated. Samples were dissolved by wet decomposition method using aqua regia and then followed by concentrated H2SO4. The solution was evaporated to dryness on a sand-bath. Study of matrix effect was carried out by comparing the absorbance of Ni and Pd with and without the adding of various concentration of metals in the samples. The calibration curve and standard addition methods were applied in the analysis of Ni and Pd. Result of the study showed that matrix effect toward absorbance of Ni is mostly caused by Al and Si contents in the concentration range of 200-500 ppm. Al content in the concentration of 500 ppm, and Si content in the concentration range of 100-500 ppm cause the matrix effect in the analysis of Pd. However, the change in absorbance of Ni and Pd is still below tolerable limit of 10%. The content of Ni in Pd-Ni / zeolit-Y, Ni-Pd/ zeolit-Y, Mo-Ni/ g-alumina, and Ni-Mo / g-alumina with calibration curve technique were 7.44 ± 0.021; 7.62 ± 0.052; 7.75 ± 0.054; and 2.69 ± 0.062 respectively and with standard addition technique were 6.82 ± 0.066; 7.23 ± 0.112; 7.39 ± 0.099; and 2.35 ± 0.055 respectively. Pd content in Pd-Ni / zeolit-Y and Ni-Pd/ zeolit-Y with calibration curve technique were 2.74 ± 0.044 and 2.49 ± 0.051 and with standard addition technique were 2.90 ± 0.015 and 2.69 ± 0.042.


Keywords


matrix effect; analysis of nickel and palladium; aqua regia; H2SO4

Full Text:

Full Text PDF


References

[1] Kojima, I., Kato, A., and Lida, C., 1992, Anal. Chim. Acta, 264

[2] Hill, A.D. Patterson, K.Y., Veillon, C., and Morris, E.R., 1986, Anal. Chim. Acta., 58

[3] Sandell, B.E. and Onishi, H. 1978, Photometric Determination of Trace of Metal General Aspect, John Wiley and Sons, New York

[4] Skoog, D.A., 1985, Principles of Instrumental Analysis, 3rd edition, Saunders College Publishing, New York

[5] Hadipranoto, N., Roto, Mudasir dan Mudjiran, 1995, Proceeding Standarisasi Metode Analisis dan Produk olahan Biji Sulfida, Puslitbang Geoteknologi-LIPI, Bandung

[6] Johnson, M.W., and Maxwell, J.A., 1981, Rock and mineral Analysis, 2nd edition, John Wiley and Sons, Inc., New York

[7] Haswell, S.J., 1988, Atomic Absorption Spectrometry; Theory and Application, Elsevier Science Publisher, Amsterdam

[8] Tunceli, A. and Turker, A.R., 2000, J. Analytical Sciences, Vol. 16

[9] Warren, H.V. and Delavault, R.E., 1959, Inst. Min, Metall., Trans., Sect. B.

[10] Olade, M. and Fletcher, K., 1974, J. Geochem. Explor., Vol. 3

[11] Wahyuni, E.T., 1998, Mempelajari Faktor-Faktor yang Berpengaruh terhadap Kestabilan Zeolit-Y sebagai Host Material Menggunakan metode Difraksi sinar-X, Seminar Jurnal Nusantara Kimia 1998, Yogyakarta

[12] Miller, J.C. and Miller, J.N., diterjemahkan Oleh Drs. Suroso, M.Sc., 1991, Statistik Untuk Kimia Analitik, Edisi kedua, Institut Teknologi Bandung, Bandung.



DOI: https://doi.org/10.22146/ijc.21914

Article Metrics

Abstract views : 1403 | views : 8272