Akuisisi Data dan Pengolahan Isyarat Elektrokardiograf Menggunakan Modul USB Dataq DI-148U

  • Henry Sulistyo Universitas Gadjah Mada
  • Thomas Sri Widodo Universitas Gadjah Mada
  • Maesadji Tjokronagoro Radiologi FK UGM/RSUP Sardjito
  • Indah Soesanti Universitas Gadjah Mada
Keywords: EKG, Akuisisi Data, Pengolahan Isyarat, Detak Jantung

Abstract

Heart signal data acquisition for the purpose of research is still an obstacle when using the standard ECG equipment from the manufacturer. The problem is the expensive equipment and the conversion of images into digital data. By using instrumentation amplifiers and electronic components in the market can be realized a simple ECG prototype by using interface Dataq DI-148U USB module for cardiac signals data acquisition.
In this study, single channel ECG prototype realized with twelve leads by manually selected. Tests using five subjects that compared the results with standard hospital ECG equipment.
From ECG prototype tests each sub section, the results are in accordance with the planning. The use of adhesive floating electrode type showed a smaller noise. IIR digital filter effectively reduces noise with SNR above of 20 dB and baseline wander at frequency cutoff 0.95 Hz. The threshold method is capable of detecting the peak R-R interval well to calculate the heart rate.

References

Ready, D.C., Biomedical Signal Processing, Tata McGraw-Hill Publishing, New Delhi, 2005.

"Low Cost, Low Power Instrumentation Amplifier AD620 data sheet," Analog Digital, Norwood, U.S.A.

Moran C. (Ed). (2000). "Building an Electrocardiogram Diagnostic System". [Online]. Available : http://cnx.org/content/col10620/1.1/

Tenedero, C. M., Raya, M. A. D., Sison, L.G, "Design and Implementation of A Single-Channel ECG Amplifier with DSP Post-Processing in Matlab", University of the Philippines. Available : http://srikrishnadevarayauniversity.academia.edu/

Chong, K.L., Holden, D., Olin, T., "Heart Rate Monitor", Analogue Electronics. ENGN3227, [Online].Available : http://users.cecs.anu.edu.au/.

Webster, J.G. (Ed), Medical Instrumentation Application and Design, 3nd Edition, John Wiley & Sons, Inc., New York, 1998.

Mortara, J. L., Morganroth, J. and Gussak, I. (Ed), "ECG Acquisition and Signal Processing", Practical Guidelines for clinical Research and Drug Development, Humana Press Inc, Totowa, NJ, p.139.

Sornmo, L., Laguna, P., "Electrocardiogram (ECG) Signal Processing", Wiley Encyclopedia of Biomedical Engineering, John Wiley & Sons, Inc, 2006

Chavan, M. S., Agarwala, R.A., Uplane, M.D, "Design of ECG Instrumentation and Implementation of Digital Filter for Noise Reduction", Recent Advances in Signal Processing, Robotics and Automation, p.36-39.

Narwaria, R. P., Verma, S., Singhal, P.K., "Removal of Baseline Wander and PowerLine Interference from ECG Signal – A Survey Approach", International Journal of Electronics Engineering, p.107-111, 3(1).2011.

Kaur, M., Singh, B., Seema, "Comparison of Different Approaches for Removal of Baseline Wander from ECG Signal", 2nd International Conference and Workshop on Emerging Trend in Technology, 2011

Gao, G. Q., "Computerised Detection and Classification of Five Cardiac Conditions", M.Eng.thesis, Auckland University of Technology, Auckland, New Zealand, 2003.

Vaseghi, S. V., Advanced Digital Signal Processing and Noise Reduction 3nd Edition, John Wiley & Sons, West Sussex, England, 2006.

Published
2014-11-10
How to Cite
Henry Sulistyo, Thomas Sri Widodo, Maesadji Tjokronagoro, & Indah Soesanti. (2014). Akuisisi Data dan Pengolahan Isyarat Elektrokardiograf Menggunakan Modul USB Dataq DI-148U. Jurnal Nasional Teknik Elektro Dan Teknologi Informasi, 2(1), 55-61. Retrieved from https://dev.journal.ugm.ac.id/v3/JNTETI/article/view/3171
Section