A COMPACT TEA CO2 LASER FOR FIELD-BASED SPECTROCHEMICAL ANALYSIS OF GEOLOGICAL SAMPLES
R Wagini(1*), K Kagawa(2)
(1) Jurusan Fisika, FMIPA UGM, Sekip Utara Bulaksumur Yogyakarta 55281
(2) Department of Physics, Faculty nof Education, University of Fukui, Fukui 910
(*) Corresponding Author
Abstract
A compact capacitor-transfer TEA CO2 laser of 180 mJ output energi and 50 ns pulse duration in half width has been designed and constructed for field use. This laser is suitable for laser microprobe spectrochemical analysis on geological and mining samples. Experiments show that the use of helium as a surrouding gas suppresses the continous emission spectrum of the plasma, thus increasing the S/N ratio. Using glass as standard samples, the minimum detectable concentrations are estimated with Zn neutral line and F ionic line to be 60 ppm and 500 ppm, respectively.
Keywords : lasers ( Compact TEA CO2 ), field-based analysis, laser microprobe analysis
Keywords : lasers ( Compact TEA CO2 ), field-based analysis, laser microprobe analysis
Keywords
lasers ( Compact TEA CO2 ), field-based analysis, laser microprobe analysis
Full Text:
PDFArticle Metrics
Abstract views : 1258 | views : 929Refbacks
- There are currently no refbacks.
ISSN 0215-9309 (Print)
Jumlah kunjungan : View my Stat.